Reference
OT Bioelettronica reference code & protocol notes
OT Bioelettronica’s example scripts define the decoding used by the native OTB
sources in myogestic/sources/otb/. These reference files remain outside the
imported and distributed package.
| File | Source | Role |
|---|---|---|
Read_muovi.m |
OTB MATLAB example v3.0 | Direct single-Muovi connection (PC = TCP server :54321). The authoritative control-byte + decode reference. |
Read_muoviAP.m |
OTB MATLAB example v2.0 | SyncStation path (PC = TCP client 192.168.76.1:54320), CRC8-framed multi-probe command strings. |
CRC8.m |
OTB MATLAB example | CRC-8 (poly 0x8C, LSB-first, init 0) used by the SyncStation / Quattrocento framed commands. |
Read_Quattrocento.m |
OTB MATLAB example v5.0 | Direct Quattrocento connection (PC = TCP client 169.254.1.10:23456). The authoritative 40-byte config + little-endian decode reference. |
FourMapsQuattrocento.m |
OTB MATLAB example | Grid/channel-map helper (maps streamed channels to electrode grids). |
ReadFromOTBiolabLightQuattrocento.m |
OTB MATLAB example | Alternative path for reading Quattrocento via the OT BioLab Light software server. MyoGestic does not use this path; it is kept for reference. |
Protocol verified against these manufacturer PDFs. They are not committed to this repository.
- Muovi probe TCP Communication Protocol v2.4
- SyncStation TCP Communication Protocol v2.8
- MuoviPro User Manual v5.1
- MuoviLite User Manual v1.1
- Quattrocento Configuration Protocol v1.7
- Quattrocento User Manual
Muovi protocol facts
- Direct Muovi: PC is TCP server on 54321; control byte
(EMG<<3) | (Mode<<1) | GO; stop =byte - 1. - Data is big-endian, 2’s-complement; EMG = int16 (2000 Hz), EEG = int24
(500 Hz). Conversion factor (gain 8) = 286.1 nV/LSB (
0.000286mV). - Frame is Fortran-order
(n_channels, samples);n_channelsdepends on(device, mode). MuoviNumChanVsMode = [38 22 38 38]. - Channel map: 1–32 bio, 33–36 IMU quaternion W/X/Y/Z, 37 buffer+trigger, 38 sample counter.
Quattrocento facts:
- PC is TCP client to 169.254.1.10:23456; little-endian int16.
- 40-byte config string, CRC-8 trailer;
ACQ_SETTbyte0 =0x80 | decim<<6 | rec_on<<5 | fsamp<<3 | nch<<1 | acq_on. - fsamp 512/2048/5120/10240 Hz; nch → 120/216/312/408 streamed channels.
- Conversion: biosignal
5/2**16/150*1000≈ 5.086e-4 mV; AUX IN5/2**16/0.5≈ 1.526e-4 V; last 8 accessory raw (counter @n-7, trigger @n-6, buffer @n-4). - The MATLAB example stops with byte0 =
0x80+ recomputed CRC. The Python source clearsacq_onwhile preserving the selected sample rate, channel count and input settings, then recomputes the CRC. Stop before reconnecting.