In isometric force tracking, the participant holds a joint still against a
transducer and matches an on-screen trapezoid through rest, ramp up, hold, ramp
down and recovery. This standard HD-EMG protocol uses a slow linear ramp and
steady plateau to measure recruitment order and discharge rates. OT Bioelettronica
provides it in its acquisition software. TrackingTask runs it in your app with
the amplifier selected by the device picker.
examples/panels/tracking_task.py runs the whole loop against a synthetic force channel if
you want to see it before wiring a load cell. This is the real version:
from myogestic import App, Streamfrom myogestic.sources import TargetSourcefrom myogestic.tracking import Trapezoidfrom myogestic.widgets import DEFAULT_DEVICES, DevicePicker, RecordButton, TrackingTask
app = App("force tracking")
emg = Stream("emg", source=DEFAULT_DEVICES[0].factory(), window_ms=200)# The target, as a second recorded stream. `TrackingTask` drives this object.target_source = TargetSource()target_stream = Stream("target", source=target_source, window_ms=200)app.streams(emg, target_stream)# Software, not hardware: there is nothing to plug in, so it attaches here. The# amplifier waits for Connect.target_stream.reconnect()
device = DevicePicker("emg")task = TrackingTask( "emg", channel=64, # the AUX channel the transducer is on trapezoid=Trapezoid(ramp_up_s=5.0, hold_s=15.0, level_pct=20.0), target=target_source,)recording = RecordButton(on_record=app.start_recording, on_stop=app.stop_recording)
@app.uidef ui(ctx): device.ui(ctx) task.ui(ctx) recording.ui(ctx)
app.run()Pass the shape to the task and construct TargetSource() without arguments.
The task forwards edits to that source, keeping the displayed and recorded
trapezoids identical.
Wire the force in
The transducer is not a biosignal channel. Connect it to an auxiliary amplifier input. Its channel follows
the bio block on the EMG stream when auxiliary inputs are enabled. Every OTB entry in the picker has an
include_aux row, off by default:
| Device | Picker row | What gets appended | Where the force lands |
|---|---|---|---|
| Muovi, Muovi+ | IMU + counters |
6: quaternion w x y z, buffer/trigger, sample counter |
no analogue input |
| Sessantaquattro+ | AUX + IMU + counters |
8: aux0, aux1, then the Muovi block |
aux0 / aux1, first two of the block |
| Quattrocento | AUX IN + accessory |
16 AUX IN, then 8 accessory | any of the 16 back-panel AUX IN |
Muovi’s accessory block contains IMU and counters, with no analogue input.
That is why its picker row names IMU and counters rather than AUX.
A plain Sessantaquattro reports a 4- or 6-wide accessory block without AUX
channels. The aux0/aux1 pair requires Sessantaquattro+. The source probes
accessory width from the ramp counter on connection. You cannot configure
that width; connect to find out what the device provides.
Do not count channel indices. The panel’s Channel dropdown lists the stream’s
own names; pick aux0 by name. Its index varies:
on a Sessantaquattro+ it is 64 at 64 channels monopolar but 32 in Bipolar, which
halves the bio block. channel= in the constructor is only where the dropdown
opens.
include_aux is off by default because Muovi and Sessantaquattro append raw
counts beside bio channels scaled to millivolts. A shared plot range cannot
usefully show both scales. Display bio channels in the
signal viewer and force in the tracking panel to give each a suitable range.
Calibrate with two captures
Map raw readings to percent of MVC with two captures:
- Zero captures rest with the participant relaxed, hand off the transducer.
- MVC keeps the peak over the next three seconds. Press it and push as hard as possible.
A load cell has a resting offset from preload, its weight or amplifier DC bias.
Normalizing against MVC alone makes a resting reading of 1.0 with an MVC of 3.0
read as 33 %, above a 30 % target. Calibration subtracts zero from the sample
and maximum so 30 % refers to 30 % of the participant’s voluntary range.
Both captures and the live trace use the chosen channel’s mean over
tail_ms, which defaults to 100 ms. Averaging reduces sample noise.
Units never come up. Calibration is a two-point map in whatever the channel
happens to carry, so it works identically on Sessantaquattro counts and on
Quattrocento volts. You do not need the transducer’s datasheet to run the
task; use it when reporting newtons.
The calibration row says Capture zero, then MVC until both captures exist. Zero and MVC sit beside that instruction; during MVC capture the same row shows the remaining seconds, then the two captured values. Start stays disabled until both captures and a force reading are available; its tooltip explains why. Recapture Zero between blocks whenever the participant repositions; the transducer’s offset moves with the grip. Setup controls are dimmed and locked while a block runs.
Give the force its own stream
Point the task at a stream carrying only force:
app.streams( Stream("emg", source=..., window_ms=200), Stream("force", source=..., window_ms=200), Stream("target", source=target, window_ms=200),)task = TrackingTask("force", target=target)An amplifier AUX input puts force on the EMG stream. Sharing that stream couples the panels’ connection state. Use a separate stream for independent force input.
To share a stream, enable the picker’s aux row and select the AUX Channel in the task. Both panels then share the device’s connection state.
Keep the target out of the picker’s reach while you are at it:
DevicePicker("emg", selectable=True, exclude=("target",))The target stream is the task’s own output. Offered as a device stream, one
Connect replaces the task’s TargetSource, detaching the source it writes to
and recording amplifier data under target.
Try it without a transducer
SyntheticForceSource simulates a transducer. It does not simulate a
participant following the target. You act as the participant by watching
the target and dragging effort to simulate pushing on a load cell.
from myogestic import Streamfrom myogestic.sources import SyntheticForceSource, TargetSource
target = TargetSource() # bare: `TrackingTask` owns the shapeforce = SyntheticForceSource() # 0.30 V at rest, 2.30 V at full effortapp.streams( Stream("force", source=force, window_ms=200), Stream("target", source=target, window_ms=200),)for name in ("force", "target"): app.ctx.streams[name].reconnect()SyntheticSource generates oscillating signals: sine by default, or stochastic
signals with waveform="emg". Use SyntheticForceSource here for a steady resting
offset and a force level that responds to effort.
Its offset and smoothing model transducer readings:
- The resting reading is not zero. A transducer has an offset, and a calibration that assumes otherwise puts every target at the wrong force. Here you can make that mistake and watch it happen.
- It smooths your slider. Dragging a slider is a step; a muscle is not.
lag_sis a first-order time constant so a drag reads as a contraction. Set it to0to follow the slider exactly.
Nothing follows the target automatically. Move effort yourself to produce
the tracking error the task measures. Automatic tracking would fabricate an
error that no participant produced.
To capture an MVC, raise effort while no block is running. The target asks
for nothing then; effort alone drives the force channel.
Shape the trapezoid
The target row shows a small trapezoid and its durations, level and repetition
count, such as 2·3·5·3·2 s · 40 % · 5 reps. Its settings icon opens the five
durations, Level, Reps and Look ahead. Captions sit above their controls
in two columns, falling back to one in narrow cells. The segment fields on
Trapezoid are ordered as they happen:
| Field | Default | What it is |
|---|---|---|
rest_s |
3.0 | Baseline before the ramp. Every analysis needs this pre-contraction reference. |
ramp_up_s |
5.0 | Linear rise to level_pct. Recruitment happens here, so slow enough to resolve it. |
hold_s |
10.0 | Seconds at the plateau. The steady-state segment discharge-rate estimates come from. |
ramp_down_s |
5.0 | Linear fall to baseline for de-recruitment, the counterpart to ramp up. |
recover_s |
5.0 | Baseline before the next repetition for fatigue control. Do not treat it as padding. |
level_pct |
30.0 | Plateau height, in percent of MVC. |
reps |
1 | How many times the shape repeats back to back. |
The sketch previews one repetition using the same corners as the plot. Its height
is normalized so the time proportions stay legible at small sizes; Level states
the plateau height. A mono summary below the plot gives the whole block’s duration,
for example 3 x 28.0 s = 84.0 s. Long row readouts ellipsize in narrow cells; hover
to read the full text.
What the participant sees
The plot frames one repetition at a time, keeping the force-target gap visible
even with five reps. The trace restarts on each repetition. The running row counts repetitions and names the
phase, such as Rep 2 of 5 · hold, above the force and signed error,
61 % MVC · error +3 %. Error is measured force minus the current target.
Stop stays beside this readout; the setup above it is dimmed and locked.
The Target and Force legend chips toggle visibility without discarding data.
Force is drawn up to the current moment. Look ahead sets how many seconds of future target are visible. A shorter preview reduces the opportunity to memorize the shape. The task measures following, so showing the whole ending from the first frame changes how the participant can approach it. Set it to zero to hide the future target or to a full repetition to show the whole trajectory in advance.
Look ahead changes the display only. It is absent from Trapezoid and
the block’s record; the commanded trajectory stays the same.
Reps roll straight on: the recover segment is the rest between them, so the block is one continuous recording and only the plot reframes.
Any segment may be zero. Zero-length segments are skipped: hold_s=0 is a legal triangle, rest_s=0 starts the ramp immediately.
Trapezoid is frozen and validates on construction, and the panel’s rows clamp to
what it accepts, so dragging a field can change the shape but never raise.
duration is one repetition, total_duration the whole block. If every duration
is zero, Start asks you to set at least one above zero before running.
At total_duration, the block ends as if Stop were pressed. The trace stays
on the plot for review.
The y range is pinned to the target so errors of 5 % and 50 % remain visibly different. Per-frame autoscaling would hide this difference.
The target is recorded as a stream
TargetSource implements the same connect / read / disconnect contract as
an amplifier, so a Stream records it exactly as it records EMG. Two channels, at
100 Hz by default:
| Channel | Name | Contents |
|---|---|---|
| 0 | target_pct |
The commanded level, percent of MVC. |
| 1 | phase |
Segment code stored as a float. This stream carries floats, so segment names are encoded as numbers. |
myogestic.sources.target.PHASE_CODES is frozen. Only add codes; never
renumber existing ones. Analysis scripts must read recordings across versions.
rest 0 · ramp_up 1 · hold 2 · ramp_down 3 · recover 4 · done 5 · idle 6idle marks time before the first Start and between blocks. Both idle and rest hold the level at 0. Separate codes keep setup time
out of the block’s rest phase, so a configured 3 s rest excludes the wait
for the operator to press Start.
The recorded trajectory shares the EMG’s local_clock() domain sample by sample.
It preserves the target used during the take without reconstructing it from
a start time and saved settings. Reconstruction would depend on two clocks
agreeing and on nobody editing a segment between noting the settings and
recording the take.
Two consequences of it being a real stream:
- Call
target_stream.reconnect().start_recordingsilently skips streams whoseinfo is Nonebecause it has no schema to allocate their arrays. An unattached target stream is absent from the session. - It emits baseline while stopped, marked
idle, to keep data continuous between blocks. A source that stops producing leaves a hole exactly where the operator was setting the next block up, and a hole is indistinguishable from a dropout at analysis time.
Recording and the task are separate buttons on purpose. Press Record, then run as many blocks as the protocol requires, then Stop. One session contains every repetition, marked by the phase channel.
Read it back
The target is a stream in the archive like any other, so everything in Record and replay applies unchanged:
from myogestic.session import open_session_store
sess = open_session_store("data/recordings/20260517-1423.mg.zip")
emg, emg_ts = sess.get_continuous("emg") # (N, n_ch) sample-majortarget, target_ts = sess.get_continuous("target") # (M, 2)
level = target[:, 0] # percent of MVCphase = target[:, 1] # PHASE_CODESUse the phase code to extract blocks from a session. The gap codes supply everything needed; no start time, settings or second clock is required.
import numpy as np
from myogestic.sources.target import PHASE_CODES
running = phase != PHASE_CODES["idle"]edges = np.flatnonzero(np.diff(running.astype(np.int8)))bounds = np.concatenate(([0], edges + 1, [len(running)]))blocks = [ (target_ts[a], target_ts[b - 1]) for a, b in zip(bounds[:-1], bounds[1:]) if b > a and running[a]]Align by timestamps because the target runs at 100 Hz and EMG at kilohertz.
Do not align by sample index. For training,
iter_aligned_windows does that for
you: each EMG window is paired with the target sample nearest its midpoint.
from myogestic.recipes.features import rmsfrom myogestic.session import iter_aligned_windows
X, Y = [], []for window, targets, ts in iter_aligned_windows( paths=["data/recordings/20260517-1423.mg.zip"], primary_stream_name="emg", aligned_stream_names=["target"], window_ms=200, hop_ms=50,): level, phase = targets["target"] if phase != 2.0: # plateau only — the ramps are a different problem continue X.append(rms(window)) Y.append(level)The exact == 2.0 holds because n_alignment_samples defaults to 1, so the phase
comes from one recorded sample rather than a mean across a segment boundary. Raise it and
compare with a tolerance, or round.
Force regression usually needs hold segments rather than ramps. Use the recorded phase channel to select them; the recording answers which samples were holds without reconstructing them from remembered settings.
See also
- Pick a device from the UI - the picker, and the
include_auxrow on every OTB entry. - Connect OTB devices - channel geometry and scaling per amplifier, including what is in each accessory block.
- Enable on-disk recording - what a session is and how it is written.
- Record and replay - reading sessions back, and the window iterators.
- Record for proportional control - the other
recorded-target protocol, on the same
TargetSourcemachinery: a signed cursor in[-1, +1]for decoder training. Force trapezoids use %MVC for motor-unit analysis.