Connect OT Bioelettronica devices

Use OTB Muovi/Muovi+ and Quattrocento directly as a Source in a Stream. They need neither Qt nor an external bridge.

Muovi / Muovi+ (Wi-Fi)

The PC is the TCP server; the probe connects to it.

  1. Hold the probe button ~5 s → it becomes a Wi-Fi access point MVxxx-ID.
  2. Join that network from the PC.
  3. Python
    from myogestic import Stream
    from myogestic.sources.otb import MuoviSource
    stream = Stream("emg", source=MuoviSource(plus=False, emg=True, mode=0),
    window_ms=1000)
    if not stream.reconnect():
    raise RuntimeError(stream.last_error)
    stream.start()

Defaults: 32-ch (Muovi) monopolar gain-8 EMG @ 2000 Hz, biosignal-only (286.1 nV/LSB → mV). Pass plus=True for 64-ch Muovi+, emg=False for EEG (500 Hz, 24-bit), include_aux=True to also stream IMU/buffer/counter.

Sessantaquattro / Sessantaquattro+ (Wi-Fi)

The PC is the TCP server on port 45454; the device connects in.

  1. Join the device’s access point, or set the PC’s IP as “Server IP address” on the sessantaquattro’s internal web page.
  2. Python
    from myogestic import Stream
    from myogestic.sources.otb import SessantaquattroSource
    stream = Stream("emg", source=SessantaquattroSource(nch_mode=3, fs_mode=2),
    window_ms=1000)
    if not stream.reconnect():
    raise RuntimeError(stream.last_error)
    stream.start()

Defaults: 64-ch monopolar gain-8 EMG @ 2000 Hz, 16-bit, biosignal-only (286.1 nV/LSB → mV). nch_mode 0..3 → 8/16/32/64 bioelectrical channels; fs_mode 0..3 → 500/1000/2000/4000 Hz. mode="bipolar" halves the channel count, mode="differential" does not, and mode="accelerometer" pins it to 8 at four times the rate. high_res=True switches to 24-bit. include_aux=True appends 2 AUX channels, the IMU quaternion, buffer/trigger and sample counter. Divide quaternion values by 16384 for a unit quaternion.

The source probes the accessory-channel count. Both protocol documents state “2 AUX + 2 accessory” for every NCH value, but a Sessantaquattro+ transmits eight. Since the frame is a flat run of words, a wrong count silently de-interleaves samples into the wrong channels. The last accessory channel is a ramp counter, so the source tries each candidate width on the first 32 samples and keeps the one that yields a clean ramp, raising if none does.

The counter also detects loss. source.dropped_samples and source.dropout_events report samples the device produced that never arrived. The first gap is logged as a warning. Check both counters after a session; sample loss may be invisible in the traces.

Quattrocento (Ethernet)

The PC is the TCP client to the amplifier (default 169.254.1.10:23456). Give the PC NIC a 169.254.x.x address on that segment.

Python
from myogestic import Stream
from myogestic.sources.otb import QuattrocentoSource
stream = Stream(
"emg", source=QuattrocentoSource(fs_mode=1, nch_mode=1), window_ms=1000
) # 2048 Hz, 192 biosignal channels (216 on the wire)
if not stream.reconnect():
raise RuntimeError(stream.last_error)
stream.start()

nch_mode 0..3 → 120/216/312/408 streamed channels; fs_mode 0..3 → 512/2048/5120/10240 Hz. Biosignal-only by default exposes the grid channels (96/192/288/384 for nch_mode 0..3) scaled to mV; include_aux=True also appends the 16 AUX IN (analog, scaled to V) and the 8 accessory channels (counter / trigger / buffer, raw). Always stop the stream before reconnecting.

Protocol references: docs/reference/otb/.