MyoGestic talks to OTB Muovi/Muovi+ and Quattrocento natively — no Qt, no external bridge. Each device is a Source you drop into a Stream.

Muovi / Muovi+ (Wi-Fi)

The PC is the TCP server; the probe connects to it.

  1. Hold the probe button ~5 s → it becomes a Wi-Fi access point MVxxx-ID.
  2. Join that network from the PC.
  3. python from myogestic import Stream from myogestic.sources.otb import MuoviSource stream = Stream("emg", source=MuoviSource(plus=False, emg=True, mode=0), window_ms=1000) stream.start()

Defaults: 32-ch (Muovi) monopolar gain-8 EMG @ 2000 Hz, biosignal-only (286.1 nV/LSB → mV). Pass plus=True for 64-ch Muovi+, emg=False for EEG (500 Hz, 24-bit), include_aux=True to also stream IMU/buffer/counter.

Sessantaquattro / Sessantaquattro+ (Wi-Fi)

The PC is the TCP server on port 45454; the device connects in.

  1. Join the device's access point, or set the PC's IP as "Server IP address" on the sessantaquattro's internal web page.
  2. python from myogestic import Stream from myogestic.sources.otb import SessantaquattroSource stream = Stream("emg", source=SessantaquattroSource(nch_mode=3, fs_mode=2), window_ms=1000) stream.start()

Defaults: 64-ch monopolar gain-8 EMG @ 2000 Hz, 16-bit, biosignal-only (286.1 nV/LSB → mV). nch_mode 0..3 → 8/16/32/64 bioelectrical channels; fs_mode 0..3 → 500/1000/2000/4000 Hz. mode="bipolar" halves the channel count, mode="differential" does not, and mode="accelerometer" pins it to 8 at four times the rate. high_res=True switches to 24-bit. include_aux=True appends the accessory channels — 2 AUX, the IMU quaternion (divide by 16384 for a unit quaternion), buffer/trigger, and the sample counter.

The accessory-channel count is probed, not configured. Both protocol documents state "2 AUX + 2 accessory" for every NCH value, but a Sessantaquattro+ transmits eight. Since the frame is a flat run of words, a wrong count does not fail — it silently de-interleaves samples into the wrong channels. The last accessory channel is a ramp counter, so the source tries each candidate width on the first 32 samples and keeps the one that yields a clean ramp, raising if none does.

That same counter measures loss: source.dropped_samples and source.dropout_events report samples the device produced that never arrived, and the first gap is logged as a warning. Check them after a session — a recording that silently lost half its samples looks completely normal otherwise.

Quattrocento (Ethernet)

The PC is the TCP client to the amplifier (default 169.254.1.10:23456). Give the PC NIC a 169.254.x.x address on that segment.

from myogestic.sources.otb import QuattrocentoSource
stream = Stream("emg", source=QuattrocentoSource(fs_mode=1, nch_mode=1),
                window_ms=1000)  # 2048 Hz, 216 streamed ch
stream.start()

nch_mode 0..3 → 120/216/312/408 streamed channels; fs_mode 0..3 → 512/2048/5120/10240 Hz. Biosignal-only by default exposes the grid channels (96/192/288/384 for nch_mode 0..3) scaled to mV; include_aux=True also appends the 16 AUX IN (analog, scaled to V) and the 8 accessory channels (counter / trigger / buffer, raw). Always stop the stream before reconnecting.

Protocol references: docs/reference/otb/.